NATIONAL B.O.D INCUBATOR
INTRODUCTIONNATIONAL B.O.D. INCUBATOR Construction Of Stainless Steel Sandwich Type 75 MM Thick PUF- filled Body. PUF Insulation Density: - 40kg/Meter³, Inner MOC of Chamber: - S.S.304/316, Outer MOC of Chamber- S.S. 304 & Wire Meshed type S.S. Trays are Provided (MOC:- S.S.304/S.S. 316).Double walled PUF- filled S.S. Door is provided along with Silicon Gasket for air tight sealing. Mechanical lock & Key provision is provided for Door. Master Sensor for temperature is provided for Controlling of chamber. The Control Panel, cooling system, & Air circulation blower are provided with attention of all standard parameter. Mounted on heavy castor wheels for easy movement
TECHNICAL SPECIFICATION OF B.O.D. INCUBATORTemperature range: Ambient +5.0°C to
60.0°C
Temperature control accuracy ±0.5°C of set
point
Temperature uniformity ±0.5°C
PUF Insulation Density:- 40kg/Meter³
model OPTIONAL SPARE ACCESSORIES
Pressure gauge/Compound gauge
Water level glass tube
Temperature Gauge
Non return valve
Moisture trap
Jointless Gasket :– Neoprene/silicon food fade rubber, Heating elements (Flange Type) Rating 6.0 kw,9.0,kw,12.0,kw
Control type: Time proportionate digital /
Microprocessor PID, Auto tune
Temperature display: 3½ digit LED
With motorized fan blower for air circulation
Inner full length acrylic door
Input voltage: 230Volts AC, 50 Hz.
To provide Calibration/ Validation certificate
- Outer cabinet mild steel powder coated & SS 304 grade dull matt finish.
- Inner chamber made of stainless steel 316/ 304 grade.
- Inner chamber stainless steel mirror finish.
- To provide all documents With DQ, IQ, OQ documentation.
Wx D x H mm | Liters | Load kw | No. Of tray |
450 x 450 x 450mm | 91 | 2.000 | 2 to 3 |
610 x 610 x 610mm | 227 | 2.000 | 2 to 4 |
600 x 650 x 900mm | 351 | 3.000 | 4 to 6 |
600 x 650 x 1200mm | 468 | 3.000 | 6 to 8 |
680 x 680 x 1300mm | 600 | 3.000 | 8 to 10 |
800 x 800 x 1250mm | 800 | 3.000 | 8 to 10 |
- PLC based system : PLC with HMI 3.7" touch screen
- Magnetic door lock feature.
- GSM system: alarms sent to predefined mobile numbers.
- Controller with printer interface to connect to EPSON Dot matrix line printer. Date, time and temperature.
- Micro processor based PID controller in lieu of above digital temperature. Controller with Auto tune facility
- Digital timer of 999 minutes with automatic heater cut-off for timed cycle.
- High temperature safety cut-off with audio/visual alarm.
Temperature Mapping Incubator: Performance qualification with temperature mapping on site. Performance validation test consist of one cycle at any one-temperature point for 2 hours period. This is done to demonstrate the uniformity of temperature within the chamber is acceptable and that the items within chamber receive a uniform temperature condition. Data is collected in printed format from up to six different points within the chamber. Temperature uniformity is considered acceptable if the deviation is less ±1ºC of set temperature.
NOTE: Every NATIONAL B.O.D. INCUBATOR is validated prior to dispatch at the factory. Those customers who do not require on site mapping may request for FREE copy of factory mapping cycle data sheet.